05-05-2011
Because the use of another manufacturer's temperature probe to calibrate the Alcotest machine does not alone compel exclusion of test results, we reverse the contrary finding of the Law Division judge in the Holland matter. Of course, the fact that the Alcotest results are not rendered automatically inadmissible thereby does not end the inquiry. Although the Alcotest machine has been found to be generally reliable, the State still bears the burden of demonstrating the "proper working order" of the device. As noted, this is accomplished by introducing into evidence the three core foundational documents, Chun, supra, 194 N.J. at 154, which the State had done in both the Holland and Pizzo matters. In other words, the State may meet its initial burden to support the admissibility of Alcotest results without reference to the calibration of the temperature probe. Thus, once the State has introduced the core documents into evidence and produced the other foundational documents in discovery, the burden of production shifts to the defendant to show why the machine was not in working order, namely, apropos to the present matters, whether and how the differences in the temperature probes had any impact at all. Absent further evidence in this regard, we discern no impediment to the admission of the Alcotest results as the State would have satisfied its ultimate burden of persuasion as to the device's reliability and accuracy.
Based on the foundational document itself, Holland raised sufficient questions as to the reliability of the Control Company's probe to warrant further inquiry. Specifically, the device was calibrated on May 26, 2009, by a State Police coordinator using a Control Company temperature probe with a serial number DDXAP2-149. During discovery, the State provided the Control Company Traceable Certificate of Calibration for Digital Thermometer as a foundational document, demonstrating the reliability of the temperature probe. However, contained on this certification are numerous serial numbers and "due dates," including a March 6, 2009 due date for temperature probe serial number 149. Although unexplained, the due date may possibly be the date the probe is due for re-certification.